For performance of ISBT Method 10.0 THC content of bev-grade CO2 and other gas applications. The Model 9000 Total Hydrocarbon Analyzer (FID) System with Automatic Calibration Control.
Insturment Requirements
Power:
Fuel Gases:
Basic CO2 software method is pre-programmed by Airborne Labs International as default. Certified laboratory back-up testing also available for result confirmation.
Detector: | Flame Ionization Detector (FID) |
Ranges: | User definable based upon calibration within: MDQ 0.01ppm, full-scale 1-200 ppm (Methane) MDQ 0.1ppm, full-scale 1-2,000ppm (Methane) MDQ 0.3ppm, full-scale 1-20,000ppm (Methane) MDQ 0.003%, full-scale 0.01-100% (Methane) |
Repeatability: | ± 1% full-scale response |
Drift: | ± 1% of full-scale over 24 hours |
Response Time: | < 5 seconds to 90% of final reading |
Alarms: | Multilevel concentration and fault alarms that result in audible and visual alarms. Alarms may also be mapped to relays to control external equipment |
Sampling: | Internal single or multipoint modules, for pre-filtered (<0.1 microns) non-condensing samples, with or without sample pump |
Optional Accessories: