Turn-Key Trace O2 Analyzer System (AII). 0-10, 0-100, 0-1000 ppm O2 in CO2 Range. 19" Rack Mount Panel Configuration.
Accuracy: | < ± 1% of full scale under constant conditions |
Analysis: | 0 - 1, 0 - 10, 0 - 100, 0 - 1000 ppm full scale ranges standard; auto-ranging or manually lock on single range |
Application: | Oxygen analysis in 100% CO2 |
Alarms: | 2 adjustable form C relay contacts non-latching; "weak sensor" indicator; power failure; system failure |
Calibration: | Certified gas of oxygen, balance nitrogen |
Compensation: | Barometric pressure and temperature; temperature controlled heated sample system |
Connections: | 1/4" compression tube fittings |
Controls: | Water resistant keypad; menu driven range selection, calibration, alarm and system functions |
Data Acquisition: | Selectable data point intervals |
Display: | Graphical LCD 5 x 2.75; resolution 0.001 ppm; displays real time ambient temperature and pressure |
Enclosure: | Painted aluminum 7.5" x 10.8" x 12.25" panel mount |
Flow Sensitivity: | None between 1 - 3 SCFH, 1 SCFH recommended |
Linearity: | > 0.995 over all ranges |
Pressure: | Inlet - regulate to 20 - 50 psig, max 150 psig; vent - atmospheric not to exceed -5" water column |
Power: | Universal; specify 110 or 220 VAC for heater system |
Range ID: | 1 - 5V output with 1V representing the lowest range |
Response Time: | 90% final full scale reading < 10 seconds |
Sample System: | Wetted parts: stainless steel consisting of flow control and sample / bypass valves; flow indicator |
Sensitivity: | < 0.5% of full scale range |
Sensor Life: | 36 months at 25°C and 1 atm; average O2 100 ppm |
Signal Output: | 4 - 20 mA isolated and 0 - 1V |
Temperature Range: | 5° to 45°C |
Warranty: | 12 months analyzer; 12 months sensor |