The BAM-8900 is a SilcoNert® passivated Benzene (Bz), Acetaldehyde (AA) and optional Methanol (MeOH) Analyzer.
8900 BAM | |
Analysis Time: | < 360 seconds |
Detector: | PID (High-sensitivity PID Optional) |
Column: | Capillary |
Oven Temperature: | 100°C, Nominal |
Carrier Gas: | Hydrogen, 15 cc/min |
Lower Detection Limit: | < 50 ppb Acetaldehyde, < 2 ppm Methanol, < 2 ppb Benzene |
Accuracy: | 1% of full-scale |
Precision: | 2% of measured value |
Span Drift (25 hrs.): | < 2% of full-scale |
Sample Flow Rate: | 250 - 500 cc/min, typical |
Analog Outputs: | (1) 0 - 20 mA or 4 - 20 mA loop power supplied, isolated. Selectable for: gas concentration, unintegrated detector signal. Options for up to 20 additional programmable 0 - 20 mA, 4 - 20 mA or voltage outputs: 0 - 1V, 0 - 5V, or 0 - 10V |
Digital Outputs: | RS-232, optional Local Area Network |
Relays: | (5) User programmale relays for concentration and diagnostic alarms (1A @ 30Vdc). Options for up to 32 additional relays available. |
Inputs: | Optional digital input board for 3 contact closure inputs. Supports start analysis, start calibration, and analyze calibration gas sample. |
Fuel Gases:
Basic CO2 software method is pre-programmed by Airborne Labs International as default. Certified laboratory back-up testing also available for result confirmation.
Optional Accessories: